Waters XBridge Columns Manuel d'utilisateur Page 5

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 12
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 4
[ Care and Use ManUal ]
XBridge
Columns 5
III. sCalInG up/down IsoCratIC methods
The following formulas will allow scale up or scale down, while maintaining
the same linear velocity, and provide new sample loading values:
If column i.d. and length are altered:
F
2
= F
1
(r
2
/r
1
)
2
Load
2
= Load
1
(r
2
/r
1
)
2
(L
2
/L
1
)
Injection volume
2
= Injection volume
1
(r
2
/r
1
)
2
(L
2
/L
1
)
Where: r = Radius of the column
F = Flow rate
L = Length of column
1 = Original, or reference column
2 = New column
IV. Troubleshooting
Changes in retention time, resolution, or backpressure are often due to column
contamination. See the Column Cleaning, Regeneration and Storage section of
this Care and Use Manual. Information on column troubleshooting problems
may be found in HPLC Columns Theory, Technology and Practice, U.D. Neue,
(Wiley-VCH, 1997), the Waters HPLC Troubleshooting Guide (Literature code
# 720000181EN) or visit the Waters Corporation website for information on
seminars (www.waters.com).
V. Column CleanInG, reGeneratIon and storaGe
a. Cleaning and Regeneration
Changes in peak shape, peak splitting, shoulders on the peak, shifts in reten-
tion, change in resolution or increasing backpressure may indicate contamina-
tion of the column. Flushing with a neat organic solvent, taking care not to
precipitate buffers, is usually sufficient to remove the contaminant. If the
flushing procedure does not solve the problem, purge the column using the
following cleaning and regeneration procedures.
Use the cleaning routine that matches the properties of the samples and/or
what you believe is contaminating the column (see Table 4 ). Flush columns
with 20 column volumes each of HPLC-grade solvents (e.g., 80 mL total for
4.6 x 250 mm column) listed in Table 4. Increasing mobile phase temperature to
35-55 ˚C increases cleaning efficiency. If the column performance is poor after
cleaning and regeneration, call your local Waters office for additional support.
Flush XBridge HILIC columns with 50:50 acetonitrile:water to remove polar
contaminants. If this flushing procedure does not solve the problem, purge the
column with 5:95 acetonitrile:water.
To clean polar contaminants from XBridge Amide columns, run a 25 minute
gradient from 0-100% water. Please note that as aqueous concentration
increases, backpressure will rapidly increase as well. Reduce flow rate when
operating at greater than 60% aqueous. Repeat if necessary.
Table 4: Cleaning and Regeneration Sequence or Options
* Use low organic solvent content to avoid precipitating buffers.
b. Storage
For periods longer than four days at room temperature, store the reversed-
phase XBridge columns and XBridge Amide columns in 100% acetonitrile. Im-
mediately after use with elevated temperatures and/or at pH extremes, store in
100% acetonitrile for the best column lifetime. Do not store columns in highly
aqueous (<20% organic) mobile phases, as this may promote bacterial growth.
If the mobile phase contained a buffer salt, flush the column with 10 column
vol umes of HPLC grade water (see Table 1 for common column volumes) and
replace with 100% acetonitrile for storage. Failure to perform this intermedi-
ate step could result in precipitation of the buffer salt in the column or system
when 100% acetonitrile is introduced. Run a gradient to 100% ACN in order
to flush all aqueous solvent from an XBridge Amide column prior to storage
in 100% ACN. Completely seal column to avoid evaporation and drying out
of the bed. For periods longer than four days, store XBridge HILIC columns in
95:5 acetonitrile:water. Do not store in buffered solvent. If the mobile phase
contained a buffered salt, flush the column with 10 column volumes of 95:5
acetonitrile:water (see Table 1 for common column volumes).
Polar Samples Non-polar Samples Proteinaceous Samples
1. water
1. isopropanol (or anappropriate
isopropanol/water mixture*)
Option 1: Inject repeated
aliquots of dimethyl sulfoxide
(DMSO)
2. methanol 2. tetrahydrofuran (THF) Option 2: gradient of 10% to 90%
B where:
A = 0.1% trifluoroacetic acid
(TFA) in water
B = 0.1% trifluoroacetic acid
(TFA) in acetonitrile (CH
3
CN)
3. tetrahydrofuran (THF) 3. dichloromethane
4. methanol 4. hexane
5. water
5. isopropanol
(followed by an appropriate
isopropanol/water mixture*)
Option 3: Flush column with 7M
guanidine hydrochloride, or 7M urea
6. mobile phase 6. mobile phase
Vue de la page 4
1 2 3 4 5 6 7 8 9 10 11 12

Commentaires sur ces manuels

Pas de commentaire