CONTENTS
I. INTRODUCTION
II. GETTING STARTED
a. Column Connectors
b. Column Installation
c. Column Equilibration
d. eCord Installation
e. Initial Column Efficiency Determination
f. VanGuard™ Pre-Columns
III. COLUMN USE
a. Sample Preparation
b. pH Range
c. Solvents
d. Pressure
e. Temperature
IV. COLUMN CLEANING, REGENERATING
AND STORAGE
a. Cleaning and Regeneration
b. Storage
V. INTRODUCING ECORD INTELLIGENT
CHIP TECHNOLOGY
a. Introduction
b. Installation
c. Manufacturing Information
d. Column Use Information
VI. ADDITIONAL INFORMATION
a. Tips for Maximizing ACQUITY UPLC BEH
Column Lifetimes
b. Recommended Flow Rates and Backpressures for
Reversed-Phase ACQUITY UPLC BEH Columns
c. Getting Started with ACQUITY UPLC BEH
HILIC Columns
d. Getting Started with ACQUITY UPLC BEH
Amide Columns
I. INTRODUCTION
Thank you for choosing a Waters ACQUITY UPLC
®
BEH Column.
The ACQUITY UPLC BEH packing materials were designed
specifically for use with the ACQUITY UPLC System and are
manufactured in a cGMP, ISO 9001 certified plant using ultra pure
reagents. Each batch of ACQUITY UPLC BEH material is tested
chromatographically with acidic, basic and neutral analytes and the
results are held to narrow specification ranges to assure excellent,
reproducible performance. Every column is individually tested and
a Performance Chromatogram and Certificate of Batch Analysis are
provided on the eCord™ intelligent chip.
ACQUITY UPLC Columns were designed and tested specifically for
use on the ACQUITY UPLC System. ACQUITY UPLC Columns will
exhibit maximum chromatographic performance and benefits ONLY
when used on the holistically-designed ACQUITY UPLC System since
the ACQUITY UPLC System and column were created and designed to
operate together. For these reasons, Waters cannot support the use
of ACQUITY UPLC Columns on any system other than an ACQUITY
UPLC System.
ACQUITY UPLC BEH Columns
[ CARE AND USE MANUAL ]
Vue de la page 0
1 2 3 4 5 6 ... 12 13

Résumé du contenu

Page 1 - ACQUITY UPLC BEH Columns

CONTENTSI. INTRODUCTIONII. GETTING STARTED a. Column Connectors b. Column Installation c. Column Equilibration d. eCord Installation e. Initial

Page 2

c. Getting Started With ACQUITY UPLC BEH HILIC Columns1. Because ACQUITY UPLC BEH HILIC Columns do not posses a bonded phase, the pH operating range

Page 3

Injection Solvents1. If possible, injection solvents should be 95% acetonitrile. The polar solvent (i.e., water, methanol, isopropanol) should be min

Page 4

in the sample are acceptable. 50:50 acetonitrile:water can provide satisfactory results.3. The injection solvent’s influence on peak shape should be

Page 5

7. If acetone is used in one or more mobile phases, do not use acetone as a sample diluent or needle wash solvent. Refer to injection solvents sectio

Page 6

II. GETTING STARTEDEach ACQUITY UPLC BEH Column comes with a Certificate of Analysis and Performance Test Chromatogram embedded within the eCord inte

Page 7

d. eCord InstallationThe eCord button should be attached to the side of the column heater module. The eCord button is magnetized and does not require

Page 8

II. COLUMN USETo ensure the continued high performance of ACQUITY UPLC BEH Columns, follow these guidelines: a. Sample Preparation 1. Sample impuriti

Page 9

Table 2. Buffer Recommendations for Using ACQUITY UPLC BEH Columns from pH 2 to 12 Additive/Buffer pKa Buffer range Volatility (±1 pH unit)Used for

Page 10 - [ CARE AND USE MANUAL ]

III. COLUMN CLEANING, REGENERATING AND STORAGEa. Cleaning and RegenerationChanges in peak shape, peak splitting, shoulders on the peak, shifts in r

Page 11

b. Storage For periods longer than four days at room temperature, store reversed-phase columns in 100% acetonitrile. For elevated temperature applicat

Page 12

c. Manufacturing Information d. Column Use InformationThe eCord chip provides the customer with column use data. The top of the screen identifies the

Page 13

4. Avoid (where possible):100% aqueous mobile phases (if possible).HPLC-grade bottled water.“Topping off” or adding “new” mobile phase to “old”