[ Care and Use ManUal ]
Contents
I. GETTING STARTED
a. Column Installation
b. Column Equilibration
c. Initial Column Efficiency Determination
II. COLUMN USE
a. Guard Columns
b. Sample Preparation
c. pH Range
d. Solvents
e. Pressure
f. Temperature
III. SCALING UP/DOWN ISOCRATIC METHODS
IV. TROUBLESHOOTING
V. COLUMN CLEANING, REGENERATING AND STORAGE
a. Cleaning and Regenerating
b. Storage
VI. CONNECTING THE COLUMN TO THE HPLC SYSTEM
a. Column Connectors and System Tubing Considerations
b. Bandspreading Minimization
c. Measuring System Bandspreading Volume & System Variance
d. Measuring System Volume
VII. ADDITIONAL INFORMATION
a. Use of Narrow-Bore (≤3.0 mm i.d.) Columns
b. Impact of Bandspreading Volume on 2.1 mm i.d.
Column Performance
c. Non-Optimized vs. Optimized LC/MS/MS System:
System Modification Recommendations
Thank you for choosing a SunFire
column. The SunFire packing
materials were designed to provide excellent peak shape, minimal
column bleed, and high mass loading. The SunFire packing materials
are manufactured in an ISO 9000 certified plant using ultra pure
reagents. Each batch of SunFire material is tested chromatographi-
cally with acidic, basic and neutral analytes and the results are held
to narrow specification ranges to assure excellent, reproducible
performance. Every column is individually tested and a Performance
Chromatogram is provided with each column along with the Certificate
of Batch Analysis.
SunFire Column Physical Characteristics
Chemistry C
18
C
8
Silica
USP Class No. L1 L7 L3
Particle Shape Spherical Spherical Spherical
Particle Sizes
2.5, 3.5,
5, 10 µm
2.5, 3.5,
5, 10 µm
5, 10 µm
Pore Size 100Å 100Å 100Å
Surface Area 340 m
2
/g 340 m
2
/g 340 m
2
/g
Carbon Load 16% 12% N/A
Endcapped Yes Yes N/A
ph Range 2-8 2-8 N/A
Temperature
Limits
Low pH = 50 ˚C
High pH = 40 ˚C
Low pH = 40 ˚C
High pH = 40 ˚C
N/A
sunFire Columns
Vue de la page 0
1 2 3 4 5 6 7 8 9 10

Résumé du contenu

Page 1 - [ Care and Use ManUal ]

[ Care and Use ManUal ]ContentsI. GETTING STARTED a. Column Installation b. Column Equilibration c. Initial Column Efficiency Determination II

Page 2

[ Care and Use ManUal ]Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com© 2010 Waters Co

Page 3

[ Care and Use ManUal ]2I. GETTING STARTEDEach SunFire column comes with a Performance Test Chromatogram. This Performance Test Chromatogram is specif

Page 4

[ Care and Use ManUal ]32. Purge the column with the mobile phase until you obtain a stable baseline.3. Verify that retention times and peak areas

Page 5

[ Care and Use ManUal ]4c. pH RangeThe recommended operating pH range for SunFire columns is 2 to 8. A listing of commonly used buffers and additives

Page 6

[ Care and Use ManUal ]5III. SCALING UP/DOWN ISOCRATIC METHODSThe following formulas will allow scale up or scale down, while maintaining the same lin

Page 7

[ Care and Use ManUal ]6b. StorageCompletely seal the column to avoid evaporation and drying out of the bed.Reversed-phase Columns (SunFire C18 and Su

Page 8

[ Care and Use ManUal ]7The presence of a void in the flow stream reduces column performance. This can occur if a Parker style ferrule is connected to

Page 9

[ Care and Use ManUal ]8b. Bandspreading MinimizationFigure 6 shows the influence of tubing internal diameter on system bandspreading and peak shape.

Page 10 - Sales Offices

[ Care and Use ManUal ]91. Remove column.2. Use acetonitrile as mobile phase A, and acetonitrile with 0.05 mg/mL uracil as mobile phase B (elimin