CONTENTS
I. INTRODUCTION
II. CONNECTING THE MASSPREP MICRO
DESALTING COLUMN WITH THE CHOSEN
LC SYSTEM
III. LC/MS SYSTEM CONFIGURATION TO
MINIMIZE MS SOURCE CONTAMINATION
IV. USING MASSPREP MICRO DESALTING
COLUMN TO MINIMIZE PROTEIN CARRYOVER
BETWEEN INJECTIONS
V. GUIDELINES FOR USE
VI. RECOMMENDED STORAGE
I. INTRODUCTION
Mass spectrometry is a powerful tool for intact protein analysis.
However, the quality of analyses can be compromised by salts in
the sample that interfere with ionization and/or form adducts with
the investigated proteins. It is, therefore, important to remove or
significantly minimize the introduction of extraneous salt ions into
the mass spectrometer. The use of on-line, desalting devices can
efficiently perform this function with minimal sample handling.
Waters MassPREP™ Micro Desalting Columns (2.1 x 5 mm, P/N
186004032) contain a polymer-based, reversed-phase packing
material (20 µm particle size, 1000 Å pore size) that can be used
to effectively desalt protein samples prior to mass spectrometry.
The phenyl phase material contained in these on-line devices
successfully retains proteins allowing the salts to be washed
to waste prior to elution of the desalted protein into the mass
spectrometer. With an optimized LC/MS method, cycle times as low
as 4 minutes are achievable. These on-line devices, when used with
an appropriate LC system injector and injector wash solution, can
be used for the sequential analyses of different samples.
II. CONNECTING THE MASSPREP MICRO DESALTING
COLUMN WITH THE CHOSEN LC SYSTEM
The MassPREP Micro Desalting Column can be used with an
appropriately performing LC system to perform the desalting
functions so long as proper fittings/tubings are used to make
connections. However, optimized performances are achieved
only when the cartridge is connected to a Waters ACQUITY or
nanoACQUITY UPLC system. For the use of the cartridge with
ACQUITY or nanoACQUITY, pre-assembled tubing kits (Intact Mass
Tubing Kit, p/n 205000548) are provided to connect the column to
the UPLC systems and mass spectrometer.
There are two configurations to connect the cartridge into the LC
fluidic path. The inlet of the cartridge can be directly screwed onto
the Valco
®
UPLC injection valve with no tubing connection.
MassPREP Micro Desalting Column
[ CARE AND USE MANUAL ]
Vue de la page 0
1 2 3

Résumé du contenu

Page 1 - [ CARE AND USE MANUAL ]

CONTENTS I. INTRODUCTION II. CONNECTING THE MASSPREP MICRO DESALTING COLUMN WITH THE CHOSEN LC SYSTEMIII. LC/MS SYSTEM CONFIGURATION TO

Page 2

This configuration would leave the cartridge operating at room temperature. The other configuration is to screw the cartridge into one end of the incl

Page 3

VI. RECOMMENDED STORAGEFor overnight storage, continuously flush the column with the mobile phase at slow flow (e.g., 50 µL/min for ACQUITY® System).