[ Care and Use ManUal ]
I. IntroductIon
This manual provides the user with information necessary to obtain
the maximum efficiency and life from the Waters Amino Acid Analysis
Column. This column has been specifically designed for the separation
and analysis of free amino acids. The column is a precision device.
Waters Corporation uses very exacting manufacturing procedures
to achieve a small average particle size, narrow range of particle
size distribution and stable particle structure in the column. These
procedures create the high efficiency column-to-column reproducibility
and long separation life built into each amino acid column from Waters.
However, care must be taken to maintain packing density and protect
it from contamination by foreign material. The following information
will aid you in obtaining the highest performance and longest life from
your column.
II. Background InformatIon
Although there are some hydrophobic interactions between the column
and certain amino acids, the primary mode of separation is that
of ion-exchange. The cation-exchange resin utilized has a large
number of negatively charged (sulfonate) functional groups which
are covalently bound to it. These negatively charged groups, in turn,
readily associate with positively charged cations when they are
present in solution. Before the amino acid sample is injected onto the
column, the resin is equilibrated at a low pH with buffer containing
a high concentration of sodium (Na+) ions. This places the column
in the sodium form. As amino acids enter the column in this acidic
buffer, they are strongly positively charged and readily exchange with
the loosely bound sodium ions.
As the buffer gradually changes to a higher pH during the gradient run,
each amino acid becomes less positively charged reflecting its pK value
and differentially elutes accordingly.
contents
I. INTRODUCTION
II. BACKGROUND INFORMATION
a. Elution System 1
b. Elution System 2
c. General Preparation Guide for One Liter of Eluent
d. Use of Non-aqueous Solvents
e. Backpressure
f. Sample Considerations
g. SEP-PAK
®
C
18
Cartridge Clean Up for Amino Acid Samples
h. Preparation Scheme
III. COLUMN CARE & USE CONSIDERATIONS
a. Column Care Considerations
b. Column Storage
c. Column Regeneration
d. Troubleshooting
e. Regeneration of Protein and Lipid Contaminated Column
f. Regeneration of Heavy Metal Contaminated Column
g. Available Service and Applications Information
h. Chromatographic Troubleshooting
i. Typical Separation
IV. COLUMN PERFORMANCE CHECKS
V. COLUMN END FITTINGS
a. End Fitting Removal Procedure
b. Column Hookup
c. Storage
APPENDIX A. AMINO ACID ANALYSIS COLUMN SPECIFICATIONS
amIno acId analysIs lIquId chromatography column
Vue de la page 0
1 2 3 4 5 6 7 8 9

Résumé du contenu

Page 1 - [ Care and Use ManUal ]

[ Care and Use ManUal ]I. IntroductIonThis manual provides the user with information necessary to obtain the maximum efficiency and life from the Wate

Page 2

[ Care and Use ManUal ]2Figure 1: Ion Exchange Mechanism CAUTION: Introduction of improperly prepared or contaminated buffers or samples onto the resi

Page 3

[ Care and Use ManUal ]3c. General Preparation Guide for One Liter of Eluent• Dissolvesaltsinapproximately900mlofpurifiedwater.• AdjustpHw

Page 4

[ Care and Use ManUal ]4III. column care & use consIderatIonsa. Column Care Considerations• ALWAYSfiltersamplestopreventexcessivepressureb

Page 5

[ Care and Use ManUal ]5f. Regeneration of Heavy Metal Contaminated Column• Oftenmetalscanberemovedbychelationunderbasicandacidiccondition

Page 6

[ Care and Use ManUal ]6Figure 3: Amino Acid Analysis – Special Amino Acids Variations in results as compared to the expected chromatogram can usual

Page 7

[ Care and Use ManUal ]7All peaks should be sharp and symmetrical. Backpressure should be less than 2,000 psi. If you suspect that the quality of you

Page 8

[ Care and Use ManUal ]8b. Column Hookup1. Direction of flow is indicated on the column.2. Connect one adaptor to the inlet and one adaptor to the out

Page 9

[ Care and Use ManUal ]Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com© 2007 Waters Cor