XSelect HSS XP 2.5 µm Columns 1[ CARE AND USE MANUAL ]CONTENTSI. INTRODUCTIONII. GETTING STARTED a. Column Connection b. Column Installation c
XSelect HSS XP 2.5 µm Columns 10[ CARE AND USE MANUAL ]c. Recommended Flow Rates and Anticipated Backpressures for Reversed-Phase XSelect HSS XP 2.5 µ
Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.comXSelect HSS XP 2.5 µm Columns 11[ CARE A
XSelect HSS XP 2.5 µm Columns 2[ CARE AND USE MANUAL ]II. GETTING STARTEDEach XSelect HSS XP 2.5 µm Column comes with a Certificate of Analysis and a
XSelect HSS XP 2.5 µm Columns 3[ CARE AND USE MANUAL ]System Volume4.4 %h5Figure 2: Determination of system band spread volume using 5-Sigma Method.Ta
XSelect HSS XP 2.5 µm Columns 4[ CARE AND USE MANUAL ]Table 3: Column Volumes (mL)Column Length (mm)Internal Diameter2.1 mm 3.0 mm 4.6 mm30 0.10 0.21
XSelect HSS XP 2.5 µm Columns 5[ CARE AND USE MANUAL ]Note: VanGuard Pre-Columns are shipped with a collet and ferrule that are NOT pre-swaged. This e
XSelect HSS XP 2.5 µm Columns 6[ CARE AND USE MANUAL ]Table 5: Buffer Recommendations for XSelect HSS XP 2.5 µm ColumnsAdditive or BufferpKaBuffer Ran
XSelect HSS XP 2.5 µm Columns 7[ CARE AND USE MANUAL ]b. Storage after Reversed-Phase UseFor periods longer than four days, store the XP 2.5 µm Column
XSelect HSS XP 2.5 µm Columns 8[ CARE AND USE MANUAL ]c. ManufacturingThe eCord provides the user with the Batch Certificate of Analysis and Performan
XSelect HSS XP 2.5 µm Columns 9[ CARE AND USE MANUAL ]3. Always remember to: Use an in-line filter unit (Part No. 205000343) or a VanGuard Pre